Difference: Feb2020TestbeamLog (72 vs. 73)

Revision 732020-03-10 - AtlasjSilicon

Line: 1 to 1
 
META TOPICPARENT name="FermilabTestbeamTop"

Feb 2020 Testbeam Log

Line: 9 to 9
 

備忘録

Tool関係

Changed:
<
<
LGAD Amp3 - DRS4までのcableの長さはちょうどいいが,LGAD - DRS4 (degitaizer family V1742)で繋ぐ場合は短い.もう少し長いのを準備するか,extendのものを準備するべき
>
>
LGAD Amp3 - DRS4までのcableの長さはちょうどいいが,LGAD - DRS4 (degitaizer family V1742)で繋ぐ場合は短い.もう少し長いのを準備するか,extendのものを準備するべき(嘘.ありました)
  AC power tapは全く使わなかった.

GPIBケーブル全く使わなかった

Changed:
<
<
HVは全く使わなかった.
>
>
HVは全く使わなかった.(一応テストするために使った)
  HV protection boardもう1枚あるといいかもしれない.
Line: 252 to 252
 
    • $ ./startrun >& /dev/null & sleep 1 ; FILENAME=`ls ../log/rawData_tlu_run0000*| tail -1` ; head -10 $FILENAME ; tail -f ${FILENAME]
    • $ ./stoprun
  • XpressK7
Added:
>
>
    • $ cd /home/atlasj/work/FNALtestbeam/XpressK7/Yarr-20200227
    • $ ./bin/spenBenchmarkDma (communication check with XpressK7)
    • $ ./bin/scanConsole -
LGAD
  • DRS4/DRS4converter -> ./converter.sh (raw data->root file)
  • latency scan:
 

Log

Mar10,2020

Line: 263 to 269
 
  • ROI: RD53A _full version
  • YArr Latency 245-->250
8:20
Changed:
<
<
  • before this time, maybe ROI did not work. in firmware we didnot apply coincidence of scinti and ROI.
>
>
  • before this time, maybe ROI did not work. in firmware we didnot apply coincidence of scinti and ROI.
 
  • ROI area
    • # of ROI signal against LGAD signal (ROI/LGAD)
Changed:
<
<
    • ROI overall -> 16/20, 8/12
>
>
    • ROI overall -> 16/20, 8/12
 
    • only RJ1 (mainly use) on ROI -> 0?
    • only RJ4 (same col, lower row) on ROI -> 0??

9:11

Changed:
<
<
  • hitbus of ROI setting is reflected when run (NOT when do analogscan. you shuld do run-cosmicGui. anyway, edit hitbus file->close cosmicGui->open calibGui->scan analog->close calibGui->opencosmicGui-> start run--> applied setting)
  • before data maybe ROI setting did not been reflected.
>
>
  • hitbus of ROI setting is reflected when run (NOT when do analogscan. you shuld do run-cosmicGui. anyway, edit hitbus file->close cosmicGui->open calibGui->scan analog->close calibGui->opencosmicGui-> start run--> applied setting)
  • before data maybe ROI setting did not been reflected.
 9:15
  • ROI region (average ROI pixel size=(250,50)
    • Col:0-80, Row:336-672 (0-336) in RJ1 were enable on hitbus --> see signal of both
Line: 299 to 305
 
    • Col:42-47, Row:481-506 (145-170) in RJ1 were enable on hitbus --> 4
    • Col:42-47, Row:461-486 (125-150) in RJ1 were enable on hitbus --> ?
    • Col:42-47, Row:466-491 (130-155) in RJ1 were enable on hitbus --> best Region Of Interesting
Added:
>
>
13:00
  • LGAD latency scan
    • (busy 3ms, with ROI,
    • latency 15: positive signal over than 0.2V events are 5 within 28k events (TLU227, LGAD105)
    • latency 30: positive 4 events within 12k (TLU228, LGAS106)
  • なぜか,channel-22,23,24に変更(EventLoopManager)したら信号(positive)が見えた.(DRS4側では,下のchannelの中の,上から5,6,7番目のchを使った)
 

Mar9,2020

7:30

 
This site is powered by the TWiki collaboration platform Powered by PerlCopyright © 2008-2022 by the contributing authors. All material on this collaboration platform is the property of the contributing authors.