META TOPICPARENT |
name="Nov2019DESYTestbeam" |
Log
Nov 25 ,2019
9:30 could not receive DACHS card at bld.6
(Because My name was not in the database.)
10:30 registered my name in the database at bld.6 room 208
11:30 received DACHS card
13:00 safety training(listen and sign name,no test)
14:00 setup
move last TB config file (ex.V2Sxx.json) to Sep2019_2 directory
make new connectivity config file:vendor6.json . and
install V6S04 /5. V6S05 is difficult to pass config. so exchange V6S05 to V6S03.
18:00 complete setup and checked digital/analog scan
install new version YarrEudaqProducer modified for memory leak problem. change the symbolic link of YarrEudaqProducerMulti from 201907xx to 20191126
apply voltage on V6S03 /4, the current is so high (-50V/100uA over)
the cuurent droped to 0 when apply some modules of vendor2 and vendor6 with connencting to YARR. (when extract display pot cable the problem did not occur)
try to connect PC GND and LV GND, PC GND and HV GND, but the problem did not resolved.
26 Nov, 2019
9:00 maybe the promble was due to Single Chip Card?
(1) module connect with only HV -> stable, 10(?)nA@-80V
(2) module connect with HV and DiplayPort -> unstable:often drop to 0 A
(3) module connect with HV and DisplayPort and LV power supply (ON) -> unstable
(4) module connect with HV and DisplayPort and LV (ON). LV GND and YARR GND short -> unstable
(5) module connect with HV and DisplayPort and LV (ON). HV GND and YARR GND short -> stable 20nA@-80V
add jumper pin which select short with HV and bias ring. (near HV Lemo connector), and short with HV GND and bias ring.
--> stable current !!
re-setup
##run plan##
(1) full depletion voltage is different with vendor and ITk.
(2) tune different way:
1.target threshold:1500, include ToT tuning
2.target threshold:1500, NOT including ToT tuning for disappearing living analog core pattern.
3. target threshold:1000, ???
11:00?? tune V6S02 : first target:2000, second target:1500, injection:10000, ToT:7, including ToT tuning
11:30 tune V6S04 : same parameters.
14:00 run control PC setup especcially editing config files. work DAQ PC from runControlPC.
14:30 first beam out
15:30 modify the stage for beam hitting sensor of RD53A as looking the monitor.
17:00 run start : DUT0:V6S02, DUT1:V6S04, both-130V, 1500e, ToT tuning ON, 10M events
27 Nov, 2019
9:00 stack...
10:30 tune modules with -100V, NO ToT tuning. (other parameters are same)
debug 50x50 vendor6 modules, but cannot
15:45 tune V6S02 : DiffVff =1 and V6S04 : DiffVff =1 (default Vff=50) , No ToT tuning
16:00 tune V6S02 : DiffVff =30 and V6S04 : DiffVff =30 , No ToT tuning
16:10 tune V6S02 : DiffVff =50 and V6S04 : DiffVff =50 , No ToT tuning-> default Diffvff=50 and tuned Diffvff=50 were not result
16:15 did digital scan for return default , and made new configfile->but not 1500e (in fact near 1650e)
16:30 changed DUT0 100V and DUT1=100V -> double peak...
17:00 changed DUT0 130V and DUT1=129.8V ,1000e, tune V6S02 :DiffVff=50 and V6S04 =50 , No ToT tuning
28 Nov, 2019
afternoon :DUT0(V6S02) tuned 1500e ToT and DUT1(V6S04) removed jumpers for grownding
DUT0 changed 80V -> 100V -> 130V -> 50V -> 10V -> 30V ->130V each with 5M events
DUT1 changed 130V -> 100V -> 80V -> 50V -> 30V -> 10V each with 5M events |